INDUSTRY ALERT !!! Well Head Unit Discards 50% of Signal !!!
An issue discovered in Well Tech units has
been tested and found to discard 50% of the signal data from the pitting
inspection function. The problem was discovered by OEM after thoroughly
analyzing a Well Tech 3 unit. A disconnected wire (shown in photos
below) which carries half of the bipolar pitting signal data is not
connected to any signal processing electronics or the digital chart deck
program. All defect signal which is transmitted on the disconnected wire
will be missed by the system operator, no matter how severe. This same
dangling wire has been reported in two additional Well Tech 3 units to
date.
-
News Update: This
problem has been discovered on a Well Tech 2 unit as well. The
wire in the Well Tech 2, however, was not left dangling.
Instead, the wire was physically connected into the header connector
which is attached to the digital to analog conversion card, but the
data was still ignored by the chart deck software.
-
OEM has designed and
implemented an upgrade solution for this problem. The solution
has been installed and tested in both Well Tech 2 and 3 systems.